Symptoms:  When processing prefabricated abutments on the N4, K5, S1, or S2 Machines

Details:  When processing prefabricated abutments, the milling machine will measure the dimensions of the prefabricated abutment blank


Possible Issues / Solutions:


Issue:  Incorrect prefabricated abutment blank inserted into blank holder

Solution:  Ensure correct blank is used / replace incorrect blank with correct blank

Note: The correct blank is defined by the CAD program in which the STl/job was created


Issue:  Measuring problem due to mechanical problem with collet chuck

Solution:  Perform collet chuck maintenance and ensure collet chuck is securely re-tightened


Issue:  Software Conflict

Solution:  Update DentalCAM to the latest official release